>Products
-
>Instruments
-
>Others
-
-
RBT320 Agglutination Automated Detection System
Tianlong RBT320 Agglutination Automated Detection System is designed for the detection of Brucellosis based on the Rose Bengal Test (RBT) method. Designed for automating laboratory workflow, Tianlong RBT320 integrates various functions, including automatic sample loading, sample mixing, photographing, AI image recognition, and result interpretation. With optimized efficiency, RBT320 can process 320 samples per hour and is suitable for large-scale screening of Brucella antibodies. Tianlong RBT320 is highly automated and can empower your lab with maximum efficiency and effectively avoid human errors, the use of the RBT method can also be cost-effective for you lab.
-
-
01Standardized detection based on the RBT method
With a temperature monitoring system to ensure the experiment temperature (22±4°C) and alarm for abnormality, reaction plates with 2cm in diameter, and 4 minutes reaction time, which fully meet the requirements of RBT Brucellosis detection international standard and ensure results accuracy
-
02Highly efficient to meet large-scale Brucellosis screening needs
Up to 320 samples can be processed per hour with RBT320. The automated instrument with standardized workflow can greatly drive lab efficiency for large-scale screening of Brucella antibodies and save human labor for your lab.
-
03Automated workflow and hands-free operation
RBT320 is highly automated and can complete the whole experiment process automatically, which can free professionals from repetitive work and avoid human errors in operation and in result interpretation.
-
04AI image recognition to interpret the results
With AI image recognition technology, consistent and precise results can be ensured and effectively avoids errors caused by visual inspection.
-
05Highly compatible with your needs
RBT320 can meet reaction conditions of different RBT reagents and is compatible with most common RBT reagents on the market.
Model | RBT320 |
Operation Time | 24T/4min(320T /hour) |
Sample Throughput | 320T (16T * 20) |
Sample Type | Blood serum |
Sample Loading Channels | 4 |
Compatible Sampling Tubes | -Compatible with various specifications of blood collection tubes -1.5mL/2.0mL centrifuge tubes, cryotubes, etc. |
Compatible Consumables | -Specialized 8-test PET reaction plates with 2cm reaction diameter -200μL disposable tips |
Pipetting Repeatability | 15μL~50μL: Er ≤ 1.5%, CV ≤ 1.5% ≥50μL: Er ≤1.0%, CV ≤ 1.2% |
Liquid level Detection | Capacitive/pressure sensing |
Report Function | Customized experiment report |
Auto Power-off | Automatic shutdown after UV disinfection |
Contamination Control | 1) Negative pressure system with HEPA filtration; 2) UV disinfection; 3) Droplet capture technology; 4) Fully enclosed biosafety cover protection |
Visualized Consumable Recognition | Sample loading status monitoring; Reaction plate loading monitoring; Identification and position of tips; Reagent status monitoring |
Operation Language | Chinese/English |
Operating Mode | PC software control via connection |
Data Connection | Ethernet |
Dimensions | 1250 mm x 710 mm x 940 mm (L x W x H) |
Power Supply | AC 220V, 50Hz |
RBT320 | |
---|---|
Operation Time | 24T/4min(320T /hour) |
Sample Throughput | 320T (16T * 20) |
Sample Type | Blood serum |
Sample Loading Channels | 4 |
Compatible Sampling Tubes | -Compatible with various specifications of blood collection tubes -1.5mL/2.0mL centrifuge tubes, cryotubes, etc. |
Compatible Consumables | -Specialized 8-test PET reaction plates with 2cm reaction diameter
-200μL disposable tips |
Pipetting Repeatability | 15μL~50μL: Er ≤ 1.5%, CV ≤ 1.5% ≥50μL: Er ≤1.0%, CV ≤ 1.2% |
Liquid level Detection | Capacitive/pressure sensing |
Report Function | Customized experiment report |
Auto Power-off | Automatic shutdown after UV disinfection |
Contamination Control | 1) Negative pressure system with HEPA filtration; 2) UV disinfection; 3) Droplet capture technology; 4) Fully enclosed biosafety cover protection |
Visualized Consumable Recognition | Sample loading status monitoring; Reaction plate loading monitoring; Identification and position of tips; Reagent status monitoring |
Operation Language | Chinese/English |
Operating Mode | PC software control via connection |
Data Connection | Ethernet |
Dimensions | 1250 mm x 710 mm x 940 mm (L x W x H) |
Power Supply | AC 220V, 50Hz |
Contact us